MICROPROCESSOR CONTROLLED OVEN MODELS: CE3F & CE5F 11/114861579INSTALLATION AND OPERATION MANUAL Thermcraft, Inc. · 3950 Overdale Road · Winst
MAINTENANCE Note: Disconnect the power cord from the power source before performing any service or maintenance on this unit. 7.1 Cleaning: Cleaning
TROUBLESHOOTING TEMPERATURE Temperature too high. 1/ Controller set too high-see section 6.3 2/ Controller failed on – call Customer Service. 3/ Wi
Indicated chamber temperature unstable 1/ ±0.1 may be normal. 3/ For Forced air models: is fan working? –verify movement of air in chamber. 4/ I
Door not sealing 1/ Adjust hinge blocks or twist the door. 2/ Confirm that unit has not been damaged and body is not square. 3/ Check physical cond
PARTS LIST Description 115v 220V EMI Filter 2800503 2800502 Adjustable Feet 2700512 2700512 Blower Motor 4880549 4880549 Control Knob 4450506 4450506
UNIT SPECIFICATIONS Weight Shipping Net CE3F 180 lbs. 83 lbs. CE5F 200 lbs. 93 lbs. Dimensions Exterior WxDxH (in.) Interior WxDxH (in.) CE3F 25.50
WIRING DIAGRAM CE3F CE5F
CE3F-2,CE5F-2
TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 PRECAUTIONS SECTION 5.0
RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be su
GRAPHIC SYMBOLS Your oven has been provided with a display of graphic symbols which should help in identifying the use and function of the available u
INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requ
PRECAUTIONS 4.1 The bottom surface of the chamber should not be used as a work surface. 4.2 This unit has been designed with a dampered
CONTROL PANEL OVERVIEW 5.1 Power Switch: The main power switch on the control panel (green lighted I/O) controls all power to the oven. It must be
OPERATION 6.1 Connection to Power Supply: Assure that the electrical power supply is properly configured and rated for the oven and plug t
B. Ten Minute Function: After the correct value for hours is set, push the RESET pad again. The blinking decimal will now move one digit to the rig